The RetractoFit is a retractable armature designed for 225 mm sensors in industrial applications. It allows the operator to mount and dismount sensors while the process is running. Safe sensor handling during process is guaranteed because insertion into the vessel without a sensor is impossible so is removal while in the measuring position
ATEX Approval | II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db |
---|---|
Autoclavable | No |
CIP | Yes |
Current in Air at 25 °C | 180 to 500 nA |
Diameter | 12 mm |
Drift at Room Temperature | < 1 % per week |
Electrical Connector | VP6 |
Electrode system | Silver platinum combination |
Electrolyte | Oxylyte B |
IECEx Approval | II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb; II 1/2 D ia IIIC T x °C Da/Db |
Maximum CO₂ Partial Pressure | 12 bar |
Measurement Principle | Electrochemical reduction of oxygen |
Measuring Range | 8 ppb to 40 ppm (DO) or 0.2 to 1000 mbar (pO₂) |
Membrane Cap | Optiflow |
Operating Temperature Range | 0 to 100 °C |
Oxygen Consumption | Ca. 100 ng/h in air at 25 °C |
Parameter | DO (Electrochemical) |
Polarization Voltage and Time | 0 ± 50 mV, ≥ 24 hours |
Pressure Range | 0 to 12 bar |
Process Connection | PG13,5 |
Required Flow | ≥ 0.1 m/s |
Response Time t90% | < 60 s at 25 °C |
Sensor Family | OxyGold B |
Sensor Output | Traditional: nA |
Serial Number | Yes |
Steam Sterilizable | No |
Surface Quality of Steel | Ra < 0.4 µm (N5) |
Temperature Sensor | NTC 22 kOhm |
Wetted Parts | Stainless Steel 1.4435 \nEPDM (Ethylene propylene elastomer) \nVMQ (Silicone elastomer) \nSilicone |
Length | 215 mm |